JOURNAL OF CLINICAL TRANSFUSION AND LABORATORY MEDICINE ›› 2017, Vol. 19 ›› Issue (6): 630-635.DOI: 10.3969/j.issn.1671-2587.2017.06.030

• Original Articles • Previous Articles     Next Articles

Performance Estimation of Sysmex XT-4000i in Body Fluid Mode for Counting Blood Cells in Cerebrospinal Fluid

GENG Li-li, LIU Yi-qing, WANG Ze-jun, et al   

  1. Department of Clinical Laboratory,Shandong Provincial Hospital Affiliated to Shandong University,Shandong Jinan 250021
  • Received:2017-08-02 Online:2017-12-20 Published:2017-12-13

Abstract: Objective To evaluate the performance of the body fluid (BF)mode on Sysmex XT-4000i for counting blood cells in cerebrospinal fluid (CSF). Methods One hundred and forty -two CSF specimens were collected from the inpatients in Shandong Provincial Hospital from June to December,2015. Blood cells of all the specimens were counted for method comparison between Sysmex XT-4000i and manual microscope (counting chamber and stained cytospin slides).Results Excellent correlations were found for counting white blood cells (y=1.17x+3.21,n=118,r2=0.98),mononuclear cells (y=0.81x+13.61,n=101,r2=0.88),polynuclear cells (y=0.82x+5.15,n=101,r2=0.88)and red blood cells (y = 1.01x+146.40,n = 62,r2= 0.99). Good agreement was found between levels of high fluorescent (HF)cells detected by XT-4000i and large cells (macrophages or malignant cells)found under the microscope (y = 0.54x+3.31,n= 101,r2= 0.27,P <0.000 1). The true positive rate of HF cells was 13.9% and their true negative rate was 96.9%. However,no correlation was found between eosnophils detected by XT-4000i and under the microscope (P >0.05). The true positive rate of eosnophils was 9.8% and their true negative rate was 80.0%. Good precision (CV≤30%)was got in white blood cells,mononuclear cells,polynuclear cells and red blood cells counted by XT-4000i. Both carry-over (≤0.3%)and linearity (r2=0.99)were also excellent for white blood cells and red blood cells. Conclusion The BF mode on XT-4000i may be routinely used for CSF cells count because of its excellent accuracy and precision. HF cells may have some predictive value in large cells in CSF.

Key words: Cerebrospinal fluid, Blood cell count, Hematology analyzer, Sysmex XT-4000i

CLC Number: