A Comparative Analysis of the Agreements of Three Domestic Kits in HPV Typing Tests

YU Hong, LU Qi-bin   

  1. Clinical Laboratory of Funing Hospital,Jiangsu,224400
  • Received:2018-01-11 Online:2019-06-20 Published:2019-06-17

Abstract: Objective To compare the consistence of HPV typing tests by using three kits that are domestically available. Methods A total of 82 positive samples and 10 negative samples were collected that had been detected with Liferiver kits in recent 2 months and the Results of HPV typing tests were compared for the agreements with the kits purchased from other two companies of Kaipu and Amply. Results The repeatability of the Results in the three domestic kits were 95.6 %,96.7% and 95.6%,respectively. The consistent rate was 51.0%(47/92),the partial consistent rate was 23.9%(22/92),and the inconsistent rate was 25.0%(23/92).The consistent rate of 86.4%(32/37)was seen in detections of single type infection,22.2%(10/45)in multiple infections,and the partial consistent rate found to be 48.9%(22/45). Conclusion of the Results in the three domestic kits varied,and the hospital should be fixed for HPV typing before and after treatments of the infection.

Key words: HPV, Domestic kits, Methodology comparison