临床输血与检验 ›› 2022, Vol. 24 ›› Issue (2): 188-192.DOI: 10.3969/j.issn.1671-2587.2022.02.010

• 基础研究 • 上一篇    下一篇

RhD抗原量子点荧光检测研究

杨倩, 陆华, 徐婷   

  1. 400038 重庆,陆军军医大学第一附属医院输血科(杨倩,徐婷); 重庆医科大学附属第二医院输血科(陆华)
  • 收稿日期:2021-11-08 发布日期:2022-04-12
  • 通讯作者: 徐婷,女,高级实验师,主要从事临床输血与检验研究,(E-mail)604537938@qq.com。共同通信作者:陆华,男,副主任医师,副教授,主要从事临床输血与检验研究,(E-mail)1580509532@qq.com。
  • 作者简介:杨倩(1993-),女,重庆人,检验技师,本科,主要从事临床输血与检验研究,(E-mail)771052231@qq.com。

Develop a Quantum Dots Fluorescent Detection of RhD Blood Group

YANG Qian, LU Hua, XU Ting   

  1. Department of Transfusion, Southwest Hospital, Army Medical University, Chongqing 400038
  • Received:2021-11-08 Published:2022-04-12

摘要: 目的 建立一种新型的RhD抗原检测方法。方法 采用量子点-抗体复合物和C1q-磁珠分离RhD抗原,根据反应前后的荧光强度比值 (Ratio value,R值) 对RhD抗原进行检测;对160例临床标本的RhD血型分型,确定量子点荧光检测法的检测性能。结果 当红细胞浓度处于1×1010~1×1011/L,35 min内可完成RhD血型抗原的检测;检测体系的组内和组间变异系数分别为3.86%和5.06%。通过量子点荧光检测法鉴定160例临床标本的RhD血型,与试管法相比符合率为100%;ROC曲线分析显示,当Cut-off值为1.610时,量子点荧光检测法在临床样本中具有较高的敏感性和特异性。结论 量子点荧光RhD血型检测可作为常规RhD抗原检测方法的一种有效补充。

关键词: Rh血型系统, D抗原, 量子点, 荧光检测

Abstract: Objective To develop a new method for RhD blood group typing. Methods In the quantum dots fluorescence detection, blood group D antigen was seperated by quantum dots-antibody complexes and magnetic beads, and identified according to the ratio of fluorescence intensity before and after the reaction (Ratio value). The detection performance of this method was validated by blood typing of 160 clinical specimens. Results When the RBC concentration is at 1×1010~1×1011/L the detection of RhD blood group antigen could be completed within 35 minutes. The intra- and inter-group variation coefficients of the detection system were 3.86% and 5.06%, respectively. Compared with the tube test method, the accuracy of the quantum dots fluorescence detection was 100% by the RhD blood group results of 160 clinical samples. ROC curve analysis showed that when the Cut-off value was 1.610, the quantum dot fluorescence detection method had the highest sensitivity and specificity. Conclusion Quantum dot fluorescent RhD blood group detection can be used as an effective supplement to conventional RhD antigen detection methods.

Key words: Rh blood group system, D antigen, Quantum dots, Fluorescence detection