• 中国科学论文统计源期刊
  • 中国科技核心期刊
  • 美国化学文摘(CA)来源期刊
  • 日本科学技术振兴机构数据库(JST)

临床输血与检验 ›› 2018, Vol. 20 ›› Issue (6): 612-614.DOI: 10.3969/j.issn.1671-2587.2018.06.017

• 临床输血 • 上一篇    下一篇

血站半自动生化分析仪定期维护和校准在降低血液ALT报废中的作用

李鹏, 孙国栋, 唐银海, 李英瑛   

  1. 056001 河北省邯郸市中心血站
  • 收稿日期:2018-03-23 出版日期:2018-12-20 发布日期:2018-11-12
  • 作者简介:李鹏(1983-),男,河北隆尧人,主管检验师,学士,主要从事血液质量控制相关研究工作,(E-mail)6766390@163.com。

Regular Calibration of Blood Semi-automatic Biochemical Analyzer Decreases Blood Discarding Due to False ALT Detection

LI Peng, SUN Guo-dong, TANG Yin-hai, et al   

  1. The Blood Center of Handan City,Hebei 056001
  • Received:2018-03-23 Online:2018-12-20 Published:2018-11-12
  • Contact: 孙国栋,女,主要从事血液质量管理,(Tel)0310-8017009。

摘要: 目的 通过对半自动生化分析仪的定期维护和校准,以减少血液因ALT不合格的报废率。方法 每3个月采用CLSI EP15-A指南中推荐的方法进行验证:5 d,每日一分析批,每日2个浓度水平,每个浓度水平上同一标本3次重复。如果因为质量控制程序或操作判一批为失控,剔除数据,并执行额外的批,操作者按照厂家说明书规定进行操作。结果 经过近5年的实验发现,半自动生化分析仪经过每3个月的维护和保养,ALT的报废率出现持续下降,分别为 5.15%(3 828/74 320)、2.14%(1 633/76 313)、2.22%(1 759/79 232)、1.57%(1 228/78 238)、0.68%(574/84 406)(P<0.01)。结论 采用CLSI EP15-A指南中推荐的方法对半自动生化分析仪进行定期维护和校准能够有效控制血液ALT的报废率,值得同行的借鉴和使用。

关键词: 血站, 维护和校准, ALT报废

Abstract: Objective To observe the maintenance and regular calibration of the semi-automatic biochemical analyzer in order to reduce blood discarding caused by ALT unqualified detections and reports. Methods Semi-automatic biochemical analyzer was calibrated every 3 months based on CLSI EP15-A guidelines. The tests were performed on one test per day for 5 days,2 concentrations per day and 3 repeats for each concentration of one sample. The additional analyses were performed and the data were eliminated in case that the quality was out of control. Results The frequencies of blood discarding due to unqualified ALT detections were gradually decreased to 5.15%(3 828/74 320),2.14%(1 633/76 313),2.22%(1 759/79 232),1.57%(1 228/78 238)and 0.68%(574/84 406)(P<0.01),respectively by a good maintenance and calibration of the instrument every three months. Conclusion The methods of nice maintenance and regular calibration of semi-automatic biochemical analyzer recommended by CLSI EP15-A may help reduce the blood discarding caused by ALT false detections and reports.

Key words: Blood station, Maintenance and calibration, ALT false detection