Influence of Pneumatic Tube System on Some Parameters of Whole Blood Cells

ZHONG Ning, LI Li-wei, Yu Jian, et al   

  1. Shanghai Sixth People's Hospital,Shanghai 200233
  • Received:2018-09-12 Online:2019-02-20 Published:2019-01-28

摘要: 目的 研究气动管道系统对全血的外周血细胞部分参数的影响。方法 选取库存时间5 d全血30袋,450 mL/袋,将30袋全血随机分为减震组与非减震组,每组各15袋。传输前后各取血标本应用Sysmex XS-800i自动血液分析仪进行血细胞参数检测;应用OLYMPUS显微镜观察瑞氏染色的红细胞形态。结果 传送前组全血的HB、HCT、RBC、MCV、MCH、MCHC、RDW、MPV、PCT、PLT、PDW、WBC分别与减震传送后即刻组、非减震传送后即刻组比较,差异均无统计学意义(P>0.05)。减震组与非减震组经PTS传送后全血保存6 h,随着保存温度变化,分别对Hb、RBC、MCV、MCHC、PCT、PLT参数有不同程度的影响(P<0.05)。另外,无论是否采取减震措施,应用PTS进行保存期5 d全血传送后,保存在一定的温度和时间内对红血细胞形态学影响不明显。结论 全血经PTS传送后即刻不会导致血细胞参数有明显的变化,立即输注给患者具有一定的安全性。

关键词: 气动管道系统, 全血, 血细胞, 形态学

Abstract: Objective To observe the Influence of pneumatic tube system on some parameters of whole blood cells. Method 30 bags of stored whole blood (450ml each) were selected to study, The bags of stored whole blood were randomly assigned to damping group or non-damping group, 15 bags in each group. Parameters of whole blood cells before the transportation were detected by Sysmex XS-800i automatic blood analyzer. The morphology of red blood cells stained with Methvlem blue and Eostm Y was observed by OLYMPUS microscope. Results There were no significant difference between damping group and non-damping group in parameters of whole blood cells (HB, HCT, RBC, MCV, MCH, MCHC, RDW, MPV, PCT, PLT, PDW, WBC) before and after the transportation(P>0.05). The statistical differences were only found in Hb, RBC, MCV, MCHC, PCT and PLT when the damping group and non-damping group stored for 6h after transportation (P<0.05). In addition, whether or not the damping measures are taken, the PTS has limited impact on the morphology of red blood cells in whole blood stored for 5 days when they were stored at a certain temperature and time after the transportation. Conclusions Pneumatic tube system has limited impact on parameters of whole blood. Immediate whole blood transfusion after transportation is safe in clinical blood transfusion.

Key words: Pneumatic tube system, Whole blood, Blood cell, Morphology