• 中国科学论文统计源期刊
  • 中国科技核心期刊
  • 美国化学文摘(CA)来源期刊
  • 日本科学技术振兴机构数据库(JST)

临床输血与检验 ›› 2019, Vol. 21 ›› Issue (3): 302-304.DOI: 10.3969/j.issn.1671-2587.2019.03.021

• 临床检验 • 上一篇    下一篇

小红细胞对全自动血细胞分析仪Sysmex XT-4000i测定血小板的影响

黄劲松   

  1. 230001 合肥市第一人民医院检验科
  • 收稿日期:2018-12-30 出版日期:2019-06-20 发布日期:2019-06-17
  • 作者简介:黄劲松(1981-),男,安徽合肥人,主管检验师,本科,主要从事血液检验工作,(Tel)15077907006。

Analysis of the Effect of Small Red Blood Cells on Platelet Assay by Automatic Blood Cell Analyzer Sysmex XT-4000i

HUANG Jin-song   

  1. The First People's Hospital of Hefei 230001
  • Received:2018-12-30 Online:2019-06-20 Published:2019-06-17

摘要: 目的 采用Sysmex XT4000-i全自动血细胞分析仪检测血小板,探讨小红细胞对其测定结果的影响。方法 采集2015年10月~2017年9月来本院就诊的210例患者的血液标本。根据红细胞平均体积大小(MCV)分为A、B、C组,将C组患者按照红细胞分布宽度是否在正常范围内分为2组。分别采用手工检测及仪器检测法对每组血液样本中的血小板进行测定。结果 A组与B组血液标本采用仪器检测法的血小板计数均高于手工检测法(P<; 0.05),C组血液标本的差异无统计学意义(P>; 0.05)。RDW异常组仪器检测法检测的血小板计数高于手工检测法(P<; 0.05),两种方法对检测RDW正常组血小板计数的差异无统计学意义(P>; 0.05)。结论 小红细胞可对全自动血细胞分析仪Sysmex XT-4000i检测的血小板计数产生影响。

关键词: 血小板, 小红细胞, 全自动血细胞分析仪

Abstract: Objective To investigate the effect of small red blood cells on the determination of platelets by using Sysmex XT4000-i automatic blood cell analyzer, manual detection and automatic blood cell analyzer Sysmex XT-4000i were used to measure each blood sample. Methods 210 blood samples were collected from patients treated in our hospital from October 2015 to September 2017. According to the average volume of red blood cells (MCV), patients were divided into A, B, C three groups, the C group of patients in accordance with the distribution of red blood cells in the normal range is divided into two groups were studied.The platelets in each blood sample were measured by manual detection and instrumentation. Results The platelet count of blood samples in group A and group B was significantly higher than that in manual test (P<0.05). There was no significant difference in blood samples between group A and group B (P>0.05). There was no significant difference in the platelet count between the two Methods (P>0.05). There was no significant difference between the two Methods in the RDW group (P>0.05). Conclusions mall red blood cells can affect the platelet count detected by automatic blood cell analyzer Sysmex XT-4000i.

Key words: Platelet, Small red blood cells, Automatic blood cell analyzer

中图分类号: